P/N: 523-52060-00
The ability to switch between precise dosing and fast conveying make this pump model the perfect complement to a reactor system. With a flow rate of 2.0 - 4,056 ml / min, when operated with single-channel pump heads, flows from feed-batch and chemostat modes of operation can be conveyed precisely. Rapid filling and emptying of the reactor system is also possible.
In addition to the selectable process parameters - speed, tubing diameter, flow quantity display, dosing volume, interval dosing and break times - the pump conveying characteristics are stored in the program. Thanks to the analogue and digital (RS 232) interfaces, processes can be automated.
All media is conveyed with a control accuracy of ±0.5% at a speed of 24 — 600 rpm. An optional foot switch can be connected for dosing by foot pressure.
Speed range | 24 - 600 rpm |
---|---|
Control accuracy motor | ±0.5 % |
Select direction of rotation | CW / CCW |
Flow rates single-channel pumps | 2.0 – 4,056 ml/min |
Flow rates multi-channel pumps | - |
Flow rate accuracy (water without counter pressure) | ±2 % |
Interval dosing | 0.001 - 9,999 ml in breaks 0.1 sec - 750 h |
Volume dosing | 0.001 - 9,999 ml |
Analog interface | for speed 0 - 10 V / 4 - 20 mA direction of rotation start/stop |
Digital Interface | RS 232 |
Protection class (EN 60529) | IP 54 |
With our Hei-FLOW peristaltic pumps, the high precision achieves minimum flow rate of only 0.005 ml/min.
When used correctly, no air bubbles emerge in the tube. The end of the tube must be deep enough in the container, from which the substance should be pumped.